This article reports on the selectivity experimentally obtained with a hand-made 24-fold 2D array with electrodes spaced at 120 μm in the rat peroneal nerve and extensor digitorum longum muscle. We call the device 2D, as all the electrode tips lie in the same plane. The device itself is a 3D multiple needle array. The design and construction of a 128-fold 3D array in silicon- and glass technology is briefly described, as well as the fabrication of a 2D 128-fold array in silicon- and LIGA technology. Special attention is given to efficiency; i.e., the ratio of the number of successful electrodes contacting a single motor fiber to the total number of electrodes in the device. We also discuss whether microfabrication technology will allow a further increase in the number of electrode sites, or if an alternative way of interfacing, namely employing cell cultures on electrode substrates, will lead to higher efficiencies.
Keywords :
LIGA; biocontrol; biomedical electrodes; biomedical electronics; microelectrodes; micromachining; neuromuscular stimulation; patient rehabilitation; 2D array; 3D multiple needle array; CMOS chip; LIGA technology; containment of cells; cultured probe; dielectrophoretic forces; efficiency; electrode spacing; extensor digitorum longum muscle; microfabrication technology; motor-fiber stimulation; motor-unit twitch; multielectrode arrays; neuro-electronic interfacing; node density estimation; rat peroneal nerve; rehabilitative neuromuscular control; selectivity; silicon and glass technology; single motor fiber; Electrodes; Extremities; Force control; Microelectrodes; Muscles; Neuromuscular; Optical fiber devices; Optimal control; Probes; Rats; Animals; Chromium Alloys; Electric Stimulation Therapy; Electrodes, Implanted; Equipment Design; Glass; Isometric Contraction; Microelectrodes; Motor Neurons; Muscle, Skeletal; Nerve Fibers; Peroneal Nerve; Polymethyl Methacrylate; Rats; Rats, Wistar; Recruitment, Neurophysiological; Sensory Thresholds; Silicones; Stress, Mechanical; Transducers;
fLanguage :
English
Journal_Title :
Engineering in Medicine and Biology Magazine, IEEE
Publisher :
ieee
ISSN :
0739-5175
Type :
jour
DOI :
10.1109/51.765188
Filename :
765188
Link To Document :
بازگشت