Abstract
Pinhole collimators are widely used to image small organs and animals. The sensitivity of knife-edge pinhole collimators has been previously estimated using an "effective diameter" formulation and experimentally described using a sin ∞ θ fit, where θ is the angle between the line segment from the center of the aperture to the photon source and its projection onto the plane of the aperture. An analytic form of the sensitivity of the pinhole collimator is derived in this paper. A numerical formula for predicting the sin ∞ θ form of the sensitivity is calculated from the analytic form. Experimental data are compared with the theoretical estimate and the sin ∞ θ prediction. The agreement is excellent.
Keywords
biomedical equipment; single photon emission computed tomography; SPECT; analytic determination; animals; line segment; medical diagnostic imaging; medical instrumentation; nuclear medicine; numerical formula; penetration; pinhole collimator sensitivity; plane of aperture; small organs; Animal structures; Apertures; Attenuation; Educational institutions; Gamma rays; Image segmentation; Optical collimators; Optical imaging; Single photon emission computed tomography; US Department of Energy; Tomography, Emission-Computed, Single-Photon;
fLanguage
English
Journal_Title
Medical Imaging, IEEE Transactions on
Publisher
ieee
ISSN
0278-0062
Type
jour
DOI
10.1109/42.938241
Filename
938241
Link To Document