• DocumentCode
    1517547
  • Title

    High-Resolution Photon Counting Using a Lens-Coupled EMCCD Gamma Camera

  • Author

    Soesbe, Todd C. ; Lewis, Matthew A. ; Slavine, Nikolai V. ; Richer, Edmond ; Bonte, Frederick J. ; Antich, Peter P.

  • Author_Institution
    Radiol. Sci. Grad. Program, Univ. of Texas, Dallas, TX, USA
  • Volume
    57
  • Issue
    3
  • fYear
    2010
  • fDate
    6/1/2010 12:00:00 AM
  • Firstpage
    958
  • Lastpage
    963
  • Abstract
    A lens-coupled electron multiplying charge-coupled device (EMCCD)-based gamma camera capable of performing photon counting for both 99mTc and 125I sources has been constructed. This system differs from previous EMCCD-based gamma cameras by using lens-coupling rather than fiber-optic coupling to transfer the light from the scintillating crystal to the EMCCD. The gamma camera described herein uses a micro-columnar CsI(Tl) crystal, two f/0.95 lenses, and a commercial camera containing the e2v CCD97 EMCCD that was cooled to -70°C. Acquisition of the video-rate frames from the CCD97 was performed using LabVIEW software. Real-time photon counting analysis of the individual scintillation flashes within the CCD97 frames was performed by using the LabVIEW IMAQ software module. An intrinsic resolution of 56 μm FWHM was measured by using a 25 μm slit collimator and 125I source. A single 0.5 mm diameter pinhole collimator was used for SPECT reconstruction of a mouse thyroid gland containing 100 μCi (3.7 MBq) of 125I uptake. We found that although the photopeak for 99mTc (140 keV) could be resolved, the photopeak for 125I (≈ 27 keV) could not be fully resolved due to the low optical transfer efficiency of dual lens coupling (<; 5%). Nonetheless, energy windowing for 125I sources was used to eliminate most of the background events, proving that high-resolution photon counting for low-energy sources can be achieved by using simple lens-coupling.
  • Keywords
    charge-coupled devices; collimators; gamma-ray apparatus; photon counting; single photon emission computed tomography; solid scintillation detectors; CCD97; LabVIEW IMAQ software; SPECT reconstruction; electron multiplying charge coupled device; fiber optic coupling; high resolution photon counting; lens coupled EMCCD gamma camera; mouse thyroid gland; pinhole collimator; scintillating crystal; Cameras; Collimators; Electrons; Energy resolution; Lenses; Optical coupling; Optical fiber devices; Photonic crystal fibers; Single photon emission computed tomography; Software performance; EMCCD; SPECT; gamma camera; high-resolution; high-sensitivity; pinhole collimator; preclinical; small animal;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.2010.2044421
  • Filename
    5485091