DocumentCode
1518334
Title
A new detection system with polycapillary conic collimator for high-localized analysis of X-ray fluorescence emission
Author
Fiorini, C. ; Longoni, A. ; Bjeoumikhov, A.
Author_Institution
Dipt. di Ingegneria Nucl., Politecnico di Milano, Italy
Volume
48
Issue
3
fYear
2001
fDate
6/1/2001 12:00:00 AM
Firstpage
268
Lastpage
271
Abstract
In this paper, we present a new detection system conceived for the X-ray fluorescence (XRF) analysis of the radiation emitted by very small areas on the sample. The system is based on the use of a polycapillary conic collimator which captures the XRF emitted only from a local area of the sample independently on how extended is the X-ray excitation. The use of a Peltier-cooled silicon drift detector allows to achieve high energy resolution with a very compact system configuration. In this first prototype, a spatial resolution of about 150 μm has been verified in the analysis of thin metal strips. Moreover, the results of a first X-Y scanning for elemental mapping on a sample surface are also reported
Keywords
X-ray detection; X-ray fluorescence analysis; silicon radiation detectors; 150 micron; Peltier-cooled Si drift detector; Si; X-ray fluorescence; polycapillary conic collimator; spatial resolution; Collimators; Energy resolution; Fluorescence; Prototypes; Radiation detectors; Silicon; Spatial resolution; Strips; X-ray detection; X-ray detectors;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/23.940063
Filename
940063
Link To Document