• DocumentCode
    1518334
  • Title

    A new detection system with polycapillary conic collimator for high-localized analysis of X-ray fluorescence emission

  • Author

    Fiorini, C. ; Longoni, A. ; Bjeoumikhov, A.

  • Author_Institution
    Dipt. di Ingegneria Nucl., Politecnico di Milano, Italy
  • Volume
    48
  • Issue
    3
  • fYear
    2001
  • fDate
    6/1/2001 12:00:00 AM
  • Firstpage
    268
  • Lastpage
    271
  • Abstract
    In this paper, we present a new detection system conceived for the X-ray fluorescence (XRF) analysis of the radiation emitted by very small areas on the sample. The system is based on the use of a polycapillary conic collimator which captures the XRF emitted only from a local area of the sample independently on how extended is the X-ray excitation. The use of a Peltier-cooled silicon drift detector allows to achieve high energy resolution with a very compact system configuration. In this first prototype, a spatial resolution of about 150 μm has been verified in the analysis of thin metal strips. Moreover, the results of a first X-Y scanning for elemental mapping on a sample surface are also reported
  • Keywords
    X-ray detection; X-ray fluorescence analysis; silicon radiation detectors; 150 micron; Peltier-cooled Si drift detector; Si; X-ray fluorescence; polycapillary conic collimator; spatial resolution; Collimators; Energy resolution; Fluorescence; Prototypes; Radiation detectors; Silicon; Spatial resolution; Strips; X-ray detection; X-ray detectors;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.940063
  • Filename