Title :
The effects of epimysial electrode location on phrenic nerve recruitment and the relation between tidal volume and interpulse interval
Author :
Schmit, Brian D. ; Mortimer, J. Thomas
Author_Institution :
Dept. of Biomed. Eng., Case Western Reserve Univ., Cleveland, OH, USA
fDate :
6/1/1999 12:00:00 AM
Abstract :
Electrode location is of vital importance to diaphragm pacing devices using electrodes implanted on the diaphragm. Complete phrenic nerve recruitment with a single epimysial electrode implanted on the abdominal surface of the diaphragm required placement within 1 cm of the motor point. Recruitment could be increased further using multiple electrodes, provided the electrodes were implanted on opposite sides of the phrenic nerve motor point. The location of the implanted electrode relative to the phrenic nerve motor point also affected the relation between the stimulus interpulse interval (IPI) and the measured tidal volume. Specifically, the authors found that electrodes implanted lateral to the phrenic nerve motor point had different tidal volume-IPI relations than electrodes placed anterior or posterior to the motor point. The authors concluded that properly placed epimysial electrodes are required to obtain adequate phrenic nerve recruitment for full time ventilation and knowledge of the relative location of the electrode with respect to motor point is necessary to predict the tidal volume produced by a specific IPI
Keywords :
biomedical electrodes; neuromuscular stimulation; pneumodynamics; 1 cm; diaphragm pacing devices; epimysial electrode location effects; implanted electrodes; interpulse interval; multiple electrodes; phrenic nerve motor point; phrenic nerve recruitment; stimulus interpulse interval; tidal volume; Abdomen; Biomedical engineering; Electrodes; Helium; Implants; Laparoscopes; Muscles; Recruitment; Ventilation; Volume measurement;
Journal_Title :
Rehabilitation Engineering, IEEE Transactions on