• DocumentCode
    1518521
  • Title

    The calibration of capacitors at the national physical laboratory, 1947

  • Author

    Rayner, G.H. ; Ford, L.H.

  • Volume
    95
  • Issue
    45
  • fYear
    1948
  • fDate
    6/1/1948 12:00:00 AM
  • Firstpage
    312
  • Lastpage
    318
  • Abstract
    Astbury´s modification of the Carey-Foster bridge is briefly described, with an account of the experience gained of its use since its installation in 1938. A more detailed description is then given of the precision Schering bridge in use at the National Physical Laboratory, in which particular attention has been paid to the screening of the components. The substandard air and mica capacitors, covering the range 200 ¿¿F to 1 ¿F, which are used in conjunction with the Schering bridge, are described, with the methods used for their calibration. The tendency in recent years has been to use the Schering bridge for measurements of capacitance and power factor, and to use the Astbury bridge chiefly for the verification of the substandards. Errors do not exceed ± 1 part in 10 000 on capacitance and ± 0.00002 on power factor when testing any but the smallest capacitors. The Appendix gives an account of the method used to determine the inductance of the internal connections of capacitors.
  • Keywords
    bridge instruments; capacitance measurement; capacitors;
  • fLanguage
    English
  • Journal_Title
    Electrical Engineers - Part II: Power Engineering, Journal of the Institution of
  • Publisher
    iet
  • Type

    jour

  • DOI
    10.1049/ji-2.1948.0077
  • Filename
    5296906