DocumentCode
1518703
Title
A list-mode time stamping data acquisition module prototype for coincidence measurements
Author
Khamzin, Murat K. ; Valentine, John D.
Author_Institution
Georgia Inst. of Technol., Atlanta, GA, USA
Volume
48
Issue
3
fYear
2001
fDate
6/1/2001 12:00:00 AM
Firstpage
570
Lastpage
574
Abstract
A list-mode data acquisition system has been developed for medical imaging and other applications in which list-mode data acquisition with time stamping is necessary. The list-mode approach is more costly and time-intensive than hardwired coincidence approaches, but due to scalability is more practical when there are many coincidence detectors. The list-mode data acquisition system was developed using commercially available components and will be used in a list mode for use with the energy-subtraction Compton scatter camera (ESCSC) “proof-of-principle” prototype. The results obtained with the prototype and a pair of NaI(TI) detectors are promising and are comparable to hardware-coincidence solutions. The time resolution of 6 ns has been achieved using time stamp acquisition from a pair of NaI(TI) scintillation detectors. The coincidence spectrum of 22Na is obtained using solely the list-mode techniques to further validate the method´s applicability and equivalence to the conventional hardwired approach
Keywords
analogue-digital conversion; biomedical electronics; data acquisition; diagnostic radiography; medical image processing; NaI(TI) detectors; NaI(TI) scintillation detectors; coincidence measurements; energy-subtraction Compton scatter camera; list-mode data acquisition; list-mode time stamping data acquisition module prototype; medical imaging; time stamp acquisition; Biomedical imaging; Biomedical measurements; Data acquisition; Detectors; Energy resolution; Prototypes; Pulse measurements; Scalability; Time measurement; Timing;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/23.940121
Filename
940121
Link To Document