DocumentCode :
1519027
Title :
Pressure Pulse Drive: A Control Method for the Precise Bending of Hydraulic Active Catheters
Author :
Ikuta, Koji ; Matsuda, Yoshikatsu ; Yajima, Daisuke ; Ota, Yusuke
Author_Institution :
Nagoya University , Nagoya, Japan
Volume :
17
Issue :
5
fYear :
2012
Firstpage :
876
Lastpage :
883
Abstract :
Our group has been developing a compact, active catheter that can be safely used in human blood vessels. The catheter is driven not by electronic mechanisms, but by a water pressure system with micro valves for multidegrees of freedom. The active catheter is composed of a soft, flexible, MRI-compatible material. In this paper, we propose what we call “pressure pulse drive,” a new control method to precisely bend the active segment. We design a new set of optimized valves for this method and experimentally verify the precise bending motion and simultaneous drive of the two active catheter segments.
Keywords :
Catheters; Hydraulic systems; Minimally invasive surgery; Pressure control; Stereolithography; Valves; Active catheter; hybrid microstereolithography; hydraulically driven system; minimally invasive surgery;
fLanguage :
English
Journal_Title :
Mechatronics, IEEE/ASME Transactions on
Publisher :
ieee
ISSN :
1083-4435
Type :
jour
DOI :
10.1109/TMECH.2011.2138711
Filename :
5770226
Link To Document :
بازگشت