Abstract
We discuss an international comparison of thermal noise-power measurements (GTRF-92-2), which has recently been completed under the auspices of the Consultative Committee for Electricity and Magnetism (CCEM). The noise temperatures of two solid-state sources with GPC-7 connectors were measured at 2, 4, and 12 GHz at the national laboratories in France, Germany, the United Kingdom, and the United States. Good agreement was found among the results from the different laboratories, with all results agreeing within the expanded uncertainties, which ranged from approximately 0.5% to 2.9%. The comparison was performed in accordance with the guidelines recently adopted by the Bureau International des Poids et Mesures (BIPM)
Keywords
electric noise measurement; measurement standards; power measurement; temperature measurement; thermal noise; 2 to 12 GHz; GPC-7 connector; GTRF-92-2; international comparison; radiometer; solid-state source; thermal noise power measurement; thermal noise temperature measurement; traveling standard; Connectors; Electric variables measurement; Guidelines; Job shop scheduling; Laboratories; NIST; Noise measurement; Particle measurements; Performance evaluation; Temperature measurement;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/19.769556
Filename
769556
Link To Document