• DocumentCode
    1519426
  • Title

    Contactless measurements of the internal capacitance of a Corbino ring in the quantum Hall regime

  • Author

    Jeanneret, Blaise ; Jeckelmann, Beat ; Hall, Blair D.

  • Author_Institution
    Fed.. Office of Metrol., Bern-Wabern, Switzerland
  • Volume
    48
  • Issue
    2
  • fYear
    1999
  • fDate
    4/1/1999 12:00:00 AM
  • Firstpage
    301
  • Lastpage
    304
  • Abstract
    A fully inductive technique has been used to observe the quantum Hall effect without any electrical connection to the two-dimensional electron gas. The ac response of the measuring system is described by a model taking into account the effect of the intrinsic capacitance of the Corbino ring. This internal capacitance is on the order of 100 pF/m for a 1 mm wide sample, a value which is not negligible in the field of ac quantum metrology
  • Keywords
    Corbino effect; capacitance measurement; quantum Hall effect; two-dimensional electron gas; AC quantum metrology; Corbino ring; contactless measurement; inductive technique; internal capacitance; quantum Hall effect; two-dimensional electron gas; Capacitance measurement; Coils; Electrical resistance measurement; Electrons; Hall effect; Laboratories; Magnetic field measurement; Measurement standards; Metrology; Quantum capacitance;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.769588
  • Filename
    769588