Title :
In-Orbit Measurement of the AOS (Acousto-Optical Spectrometer) Response Using Frequency Comb Signals
Author :
Mizobuchi, Satoko ; Kikuchi, Ken-ichi ; Ochiai, Satoshi ; Nishibori, Toshiyuki ; Sano, Takuki ; Tamaki, Kenta ; Ozeki, Hiroyuki
Author_Institution :
Japan Aerosp. Exploration Agency, Tsukuba, Japan
fDate :
6/1/2012 12:00:00 AM
Abstract :
The in-orbit response characteristic of the AOS (Acousto-Optical Spectrometer) was estimated from spectral profiles taken by using a frequency comb generator, which was originally used for frequency axis calibration. Frequency comb signals taken under various environmental conditions in orbit were consolidated to deduce the response profile of the spectrometer. The fluctuation bandwidth calculated from the deduced spectral response compared well with that derived from the noise characteristic of the spectrometer.
Keywords :
acousto-optical devices; geophysical equipment; remote sensing; spectrometers; acousto-optical spectrometer; fluctuation bandwidth; frequency comb generator; frequency comb signals; in-orbit measurement; noise characteristics; spectral profiles; Data models; Diode lasers; Extraterrestrial measurements; Frequency measurement; Instruments; Orbits; Radio frequency; AOS; Response function; comb generator; spectrometer;
Journal_Title :
Selected Topics in Applied Earth Observations and Remote Sensing, IEEE Journal of
DOI :
10.1109/JSTARS.2012.2196413