DocumentCode :
1520047
Title :
A set of waveguide primary thermal noise standards and related calibration systems for the frequency range 8.2-40 GHz
Author :
Achkar, Joseph
Author_Institution :
BNM-LCIE, Fontenay-aux-Roses, France
Volume :
48
Issue :
2
fYear :
1999
fDate :
4/1/1999 12:00:00 AM
Firstpage :
638
Lastpage :
641
Abstract :
The national metrology laboratory, BNM-LCIE, France, has developed a set of four new waveguide primary thermal noise standards for the frequency band from 8.2 GHz to 40 GHz. Each noise standard is made with silicon carbide microwave absorber and inserted into a section of waveguide made in platinum 10% rhodium; the whole is put inside an electrical oven. The output noise temperature range of the standards obtained from temperature and attenuation measurements is within 948-1012 K±1.21.8 K, in one standard-type uncertainty. The calibration of gas discharge noise sources and/or solid-state noise sources is achieved by comparison to standards using a switching radiometer. The excess noise ratio given is within 15.35-15.65 dB for gas discharge sources and within 14-16 dB for solid-state sources, with typical uncertainty of 0.02 dB
Keywords :
calibration; electric noise measurement; measurement standards; microwave measurement; rectangular waveguides; thermal noise; 8.2 to 40 GHz; Pt-Rh; SiC; attenuation measurement; calibration; electrical oven; electromagnetic metrology; gas discharge noise source; platinum-rhodium waveguide; primary thermal noise standard; silicon carbide microwave absorber; solid-state noise source; temperature measurement; Discharges; Frequency; Laboratories; Metrology; Microwave ovens; Platinum; Silicon carbide; Solid state circuits; Standards development; Temperature distribution;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/19.769676