• DocumentCode
    1524142
  • Title

    Low magnetic field measurement by a separated NMR detector using flowing water

  • Author

    Woo, B.C. ; Kim, C.G. ; Park, Po Gyu ; Kim, C.S. ; Shifrin, V.Y.

  • Author_Institution
    Korea Res. Inst. of Stand. & Sci., Taejon, South Korea
  • Volume
    33
  • Issue
    5
  • fYear
    1997
  • fDate
    9/1/1997 12:00:00 AM
  • Firstpage
    4345
  • Lastpage
    4348
  • Abstract
    Using polarized flowing water, NMR detection at a high field region separated far away from the measured field was investigated to estimate ranges and conditions for a wide-range magnetometer. We obtained the Lorentzian-type NMR signal with the full-half-maximum width of 320 nT and the S/N ratio of 10.5, and measured the magnetic field from several mT down to a few μT with the same enhancement of signals by a separated NMR detector
  • Keywords
    NMR spectrometers; magnetic field measurement; magnetometers; FWHM; Lorentzian signal; S/N ratio; low magnetic field measurement; magnetometer; polarized flowing water; separated NMR detector; Coils; Detectors; Magnetic field measurement; Magnetization; Magnetometers; Nuclear magnetic resonance; Polarization; Protons; Radio frequency; Time measurement;
  • fLanguage
    English
  • Journal_Title
    Magnetics, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9464
  • Type

    jour

  • DOI
    10.1109/20.620444
  • Filename
    620444