• DocumentCode
    1525536
  • Title

    An improved approximation for the spatial impulse response of a rectangular transducer

  • Author

    Teo, Tat-Jin

  • Author_Institution
    Boston Sci. Corp., San Jose, CA, USA
  • Volume
    45
  • Issue
    1
  • fYear
    1998
  • Firstpage
    76
  • Lastpage
    83
  • Abstract
    One common approximation for a field point located in the far field of the transducer is to trapezoidally-shaped function to approximate the true spatial impulse response. A quadrilaterally shaped function is proposed here. The quadrilaterally shaped function also includes the trapezoidally shaped function as a special case. This paper shows that the quadrilateral approximation applied in the specifically defined field location is superior to the trapezoidal approximation for the true spatial impulse response function when the transducer elements have the aspect ratio of approximately 10, typical of those found in 1.5 D arrays.
  • Keywords
    acoustic field; acoustic transducers; approximation theory; transient response; ultrasonic transducers; 1.5 D array; aspect ratio; far field; quadrilateral approximation; rectangular transducer; spatial impulse response; trapezoidal approximation; Apertures; Biomedical imaging; Biomedical transducers; Convolution; Fourier transforms; Geometry; Sampling methods; Ultrasonic imaging; Ultrasonic transducers;
  • fLanguage
    English
  • Journal_Title
    Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0885-3010
  • Type

    jour

  • DOI
    10.1109/58.646913
  • Filename
    646913