Abstract :
This paper reports the energy resolutions obtained from two 3-D CZT γ-ray spectrometers. Each device is a 1 cm3 cube CZT crystal with an 11×11 pixellated anode. For single-interaction events in the detector, the gamma ray interaction location can be determined in 3-D. With the volume of each detector conceptually divided into 11×11×18 voxels, the spectra of single-interaction events from each of the voxels were collected. The energy resolutions at 662 keV from a single voxel are typically 9.3 keV and 9.9 keV for the two detectors, with electronic noise contributing ~7 keV and ~6 keV respectively. A spatially-varying energy resolution has been observed for both detectors, which accounts for the degradation of energy resolution of the overall spectrum combined from all the voxels. The causes for the variations have been analyzed, and the impact of a spatially-varying energy resolution on a CZT gamma-ray spectrometer has been investigated
Keywords :
II-VI semiconductors; cadmium compounds; gamma-ray spectrometers; position sensitive particle detectors; semiconductor counters; 3D position sensitive CZT gamma-ray spectrometers; CdZnTe; electronic noise; energy resolution; gamma ray interaction location; pixellated anode; single-interaction events; spatially-varying energy resolution; voxels; Aerospace engineering; Anodes; Cathodes; Degradation; Detectors; Electron traps; Energy resolution; Event detection; Signal processing; Spectroscopy;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/23.775512
Filename :
775512
Link To Document :
بازگشت