DocumentCode :
1531195
Title :
Correction for recombination loss of ions in ionization chambers
Author :
Takata, N. ; Yin, Zaizhe ; Li, Aiwu
Author_Institution :
Electrotech. Lab., Ibaraki, Japan
Volume :
46
Issue :
3
fYear :
1999
fDate :
6/1/1999 12:00:00 AM
Firstpage :
326
Lastpage :
332
Abstract :
The signal charge loss due to initial recombination and that due to volume recombination have been measured separately using parallel plate ionization chambers irradiated with cobalt-60 gamma rays, over a wide range of applied voltage and air humidity. It is shown that both initial and volume recombination depend on the electric field strength and also on humidity through ion clustering reaction with water molecules. In spite of the usual assumptions, a linear relation has not been observed between the inverses of the signal current and applied voltage for initial recombination, nor between the inverses of the signal current and square of the applied voltage for volume recombination. Methods and data which can be used for estimation of initial and volume recombination losses of ions in ionization chambers are proposed
Keywords :
gamma-ray detection; ion recombination; ionisation chambers; air humidity; electric field strength; initial recombination; ion clustering reaction; ionization chambers; parallel plate ionization chambers; recombination loss; signal charge loss; signal current; volume recombination; water molecules; Electric potential; Electrodes; Equations; Humidity; Ionization chambers; Ionizing radiation; Laboratories; Metrology; Spontaneous emission; Voltage;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/23.775538
Filename :
775538
Link To Document :
بازگشت