• DocumentCode
    1531444
  • Title

    Electronics for a prototype variable field of view PET camera using the PMT-quadrant-sharing detector array

  • Author

    Li, Hongdi ; Wong, Wai-Hoi ; Zhang, Nan ; Wang, Junming ; Uribe, Jorge ; Baghaei, Hossain ; Yokoyama, Shigeru

  • Author_Institution
    M.D. Anderson Cancer Center, Texas Univ., Houston, TX, USA
  • Volume
    46
  • Issue
    3
  • fYear
    1999
  • fDate
    6/1/1999 12:00:00 AM
  • Firstpage
    546
  • Lastpage
    550
  • Abstract
    Electronics for a prototype high-resolution PET camera with eight position-sensitive detector modules has been developed. Each module has 16 BGO (Bi4Ge3O12) blocks (each block is composed of 49 crystals). The design goals are component and space reduction. The electronics is composed of five parts: front-end analog processing, digital position decoding, fast timing, coincidence processing and master data acquisition. The front-end analog circuit is a zone-based structure (each zone has 3×3 PMTs). Nine ADCs digitize integration signals of an active zone identified by eight trigger clusters; each cluster is composed of six photomultiplier tubes (PMTs). A trigger corresponding to a gamma ray is sent to a fast timing board to obtain a time-mark, and the nine digitized signals are passed to the position decoding board, where a real block (four PMTs) can be picked out from the zone for position decoding. Lookup tables are used for energy discrimination and to identify the gamma-hit crystal location. The coincidence board opens a 70-ns initial timing window, followed by two 20-ns true/accidental time-mark lookup table windows. The data output from the coincidence board can be acquired either in sinogram mode or in list mode with a Motorola/IRONICS VME-based system
  • Keywords
    analogue processing circuits; analogue-digital conversion; arrays; cameras; data acquisition; decoding; modules; photodetectors; photomultipliers; positron emission tomography; table lookup; timing circuits; Bi4Ge3O12; Motorola/IRONICS VME-based system; analogue-digital convertors; coincidence board; coincidence processing; component reduction; digital position decoding; energy discrimination; fast timing board; front-end analog processing; gamma-hit crystal location; gamma-ray triggering; high-resolution PET camera; initial timing window; integration signal digitization; list mode; master data acquisition; photomultiplier tube quadrant-sharing detector array; position decoding board; position-sensitive detector modules; prototype variable field-of-view PET camera electronics; sinogram mode; space reduction; time-mark; trigger clusters; true/accidental time-mark lookup table windows; zone-based analog circuit structure; Bismuth; Cameras; Crystals; Data acquisition; Decoding; Position sensitive particle detectors; Positron emission tomography; Prototypes; Table lookup; Timing;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.775577
  • Filename
    775577