Abstract :
This paper describes progress in the design and construction of a single-plane PET tomograph having a spatial resolution approaching 1 mm. The system consists of a 12-cm diameter ring with 360 LSO (Lu2SiO5) detectors viewed by 30 photomultiplier tubes. Thin (5-mm) crystals and a low energy threshold are used. Crystals are identified using both position arithmetic and energy criteria. To date, the system construction has been completed, system tuning has been carried out and imaging studies have begun
Keywords :
biological techniques; biomedical equipment; image processing equipment; image resolution; photomultipliers; positron emission tomography; zoology; 1 mm; 12 cm; 5 mm; LSO detectors; Lu2SiO5; crystal identification; energy criteria; energy threshold; imaging studies; photomultiplier tubes; position arithmetic; single-plane PET; small-animal PET imaging device; spatial resolution; system construction; system tuning; thin crystals; Animals; Circuits; Detectors; High-resolution imaging; Image resolution; Instruments; Photonic crystals; Positron emission tomography; Signal resolution; Spatial resolution;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/23.775590
Filename :
775590
Link To Document :
بازگشت