• DocumentCode
    1531529
  • Title

    Development of a small animal PET imaging device with resolution approaching 1 mm

  • Author

    Correia, J.A. ; Burnham, C.A. ; Kaufman, D. ; Fischman, A.J.

  • Author_Institution
    Massachusetts Gen. Hosp., Boston, MA, USA
  • Volume
    46
  • Issue
    3
  • fYear
    1999
  • fDate
    6/1/1999 12:00:00 AM
  • Firstpage
    631
  • Lastpage
    635
  • Abstract
    This paper describes progress in the design and construction of a single-plane PET tomograph having a spatial resolution approaching 1 mm. The system consists of a 12-cm diameter ring with 360 LSO (Lu2SiO5) detectors viewed by 30 photomultiplier tubes. Thin (5-mm) crystals and a low energy threshold are used. Crystals are identified using both position arithmetic and energy criteria. To date, the system construction has been completed, system tuning has been carried out and imaging studies have begun
  • Keywords
    biological techniques; biomedical equipment; image processing equipment; image resolution; photomultipliers; positron emission tomography; zoology; 1 mm; 12 cm; 5 mm; LSO detectors; Lu2SiO5; crystal identification; energy criteria; energy threshold; imaging studies; photomultiplier tubes; position arithmetic; single-plane PET; small-animal PET imaging device; spatial resolution; system construction; system tuning; thin crystals; Animals; Circuits; Detectors; High-resolution imaging; Image resolution; Instruments; Photonic crystals; Positron emission tomography; Signal resolution; Spatial resolution;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.775590
  • Filename
    775590