DocumentCode
1531529
Title
Development of a small animal PET imaging device with resolution approaching 1 mm
Author
Correia, J.A. ; Burnham, C.A. ; Kaufman, D. ; Fischman, A.J.
Author_Institution
Massachusetts Gen. Hosp., Boston, MA, USA
Volume
46
Issue
3
fYear
1999
fDate
6/1/1999 12:00:00 AM
Firstpage
631
Lastpage
635
Abstract
This paper describes progress in the design and construction of a single-plane PET tomograph having a spatial resolution approaching 1 mm. The system consists of a 12-cm diameter ring with 360 LSO (Lu2SiO5) detectors viewed by 30 photomultiplier tubes. Thin (5-mm) crystals and a low energy threshold are used. Crystals are identified using both position arithmetic and energy criteria. To date, the system construction has been completed, system tuning has been carried out and imaging studies have begun
Keywords
biological techniques; biomedical equipment; image processing equipment; image resolution; photomultipliers; positron emission tomography; zoology; 1 mm; 12 cm; 5 mm; LSO detectors; Lu2SiO5; crystal identification; energy criteria; energy threshold; imaging studies; photomultiplier tubes; position arithmetic; single-plane PET; small-animal PET imaging device; spatial resolution; system construction; system tuning; thin crystals; Animals; Circuits; Detectors; High-resolution imaging; Image resolution; Instruments; Photonic crystals; Positron emission tomography; Signal resolution; Spatial resolution;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/23.775590
Filename
775590
Link To Document