• DocumentCode
    1535133
  • Title

    Tachycardia recognition algorithms for implantable systems

  • Author

    Khoury, D.S. ; Wilkoff, B.L.

  • Author_Institution
    Dept. of Cardiology, Cleveland Clinic Found., OH, USA
  • Volume
    9
  • Issue
    2
  • fYear
    1990
  • fDate
    6/1/1990 12:00:00 AM
  • Firstpage
    40
  • Lastpage
    42
  • Abstract
    The detection algorithm of current antitachycardia pacing systems incorporates derivatives of heart rate as the only objective parameter used for recognition. Present pacing systems use one or more rate-related parameters: (1) rate cutoff, (2) sudden onset, (3) rate stability, and (4) sustained high rate. Recently, the development of antitachycardia systems has focused on combining the four detection criteria to form multiple recognition algorithms. Nevertheless, inability to differentiate between supraventricular and ventricular tachycardias at compatible rates remains a limitation. Incorporating atrial sensing capability, in addition to ventricular sensing capability, an alternative strategy that potentially could enhance diagnostic accuracy, is discussed. The incorporation of time and frequency analysis, arterial pressure, and stroke volume are also examined.<>
  • Keywords
    pacemakers; antitachycardia systems; atrial sensing capability; diagnostic accuracy; heart rate derivatives; implantable systems; rate cutoff; rate stability; sudden onset; supraventricular tachycardia; sustained high rate; tachycardia recognition algorithms; ventricular tachycardia; Algorithm design and analysis; Biomedical measurements; Heart rate; Heart rate detection; Hemodynamics; Interference; Particle measurements; Physiology; Rhythm; Termination of employment;
  • fLanguage
    English
  • Journal_Title
    Engineering in Medicine and Biology Magazine, IEEE
  • Publisher
    ieee
  • ISSN
    0739-5175
  • Type

    jour

  • DOI
    10.1109/51.57867
  • Filename
    57867