• DocumentCode
    1536884
  • Title

    A low-light-level sensor for medical diagnostic applications

  • Author

    Perelman, Yevgeny ; Ginosar, Ran

  • Author_Institution
    Dept. of Electr. Eng., Technion-Israel Inst. of Technol., Haifa, Israel
  • Volume
    36
  • Issue
    10
  • fYear
    2001
  • fDate
    10/1/2001 12:00:00 AM
  • Firstpage
    1553
  • Lastpage
    1558
  • Abstract
    A low-level light sensor and preprocessor for a disposable medical probe is described. A 10-Hz 4-μA/cm2 input optical signal in the visible spectrum is assumed in chemoluminant diagnostic applications. A single-bit first-order sigma-delta modulator has been employed for analog-to-digital conversion, thanks to its robustness, simplicity, inherent linearity, and high signal-to-noise ratio. A current buffer has been used to replicate the weak signal with sufficient bandwidth. Tests have shown the feasibility of this approach, and have also enabled improvements of the design
  • Keywords
    biomedical electronics; buffer circuits; chemiluminescence; patient diagnosis; photodetectors; sigma-delta modulation; analog-to-digital conversion; chemoluminant reaction; current buffer; disposable medical probe; low-light-level sensor; medical diagnostics; optical signal; preprocessor; sigma-delta modulator; signal-to-noise ratio; visible spectrum; Analog-digital conversion; Biomedical optical imaging; Delta-sigma modulation; Medical diagnosis; Medical diagnostic imaging; Optical buffering; Optical modulation; Optical sensors; Probes; Sensor phenomena and characterization;
  • fLanguage
    English
  • Journal_Title
    Solid-State Circuits, IEEE Journal of
  • Publisher
    ieee
  • ISSN
    0018-9200
  • Type

    jour

  • DOI
    10.1109/4.953484