DocumentCode :
1542137
Title :
LMIRA: list-mode iterative reconstruction algorithm for SPECT
Author :
Bouwens, L. ; Van de Walle, R. ; Gifford, H. ; King, M. ; Lemahieu, I. ; Dierckx, R.A.
Author_Institution :
ELIS Dept., Ghent Univ., Belgium
Volume :
48
Issue :
4
fYear :
2001
fDate :
8/1/2001 12:00:00 AM
Firstpage :
1364
Lastpage :
1370
Abstract :
The goal of this project is to investigate a new method for using list-mode data instead of binned projection data for single photon emission computed tomography (SPECT) reconstruction. List-mode acquisition was once standardly available for planar scintigraphy and is now becoming available for SPECT acquisition. The authors implemented maximum-likelihood iterative reconstruction with list-mode data for SPECT using a new projector/backprojector pair, which employs an intermediate stage to separate the source and camera specific components. In two dimensions, the intermediate stage is a circle surrounding the object. An emission radiance distribution is calculated at the circle from the estimate of the source distribution in the object. This leads to the probability that a photon, coming from the source distribution, leaves the circle surrounding the object at a certain point, going in a specific direction. The algorithm uses the emission radiance distribution to obtain the probabilities used in the iterative reconstruction method. The model takes into account distance-dependent resolution and nonuniform attenuation. Results illustrate how the individual steps of the algorithm work. The list-mode iterative reconstruction algorithm is compared to the standard MLEM iterative reconstruction method, which employs binned data. It is concluded that the finer sampling provided by list-mode data results in superior image quality. By using the intermediate stage, the authors obtained a structure that can very easily be adapted to different types of collimators and acquisition protocols
Keywords :
image reconstruction; iterative methods; medical image processing; single photon emission computed tomography; LMIRA; SPECT; acquisition protocols; binned projection data; camera specific components; collimators; emission radiance distribution; list-mode iterative reconstruction algorithm; maximum-likelihood iterative reconstruction; medical diagnostic imaging; nuclear medicine; planar scintigraphy; projector/backprojector pair; source specific components; Attenuation; Cameras; Image quality; Image reconstruction; Image sampling; Iterative algorithms; Iterative methods; Maximum likelihood estimation; Reconstruction algorithms; Single photon emission computed tomography;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/23.958356
Filename :
958356
Link To Document :
بازگشت