• DocumentCode
    1542309
  • Title

    Uranium pinhole collimators for 511-keV photon SPECT imaging of small volumes

  • Author

    Tenney, C.R. ; Tornai, M.P. ; Smith, M.F. ; Turkington, T.G. ; Jaszczak, R.J.

  • Author_Institution
    Duke Univ. Med. Center, Durham, NC, USA
  • Volume
    48
  • Issue
    4
  • fYear
    2001
  • fDate
    8/1/2001 12:00:00 AM
  • Firstpage
    1483
  • Lastpage
    1489
  • Abstract
    Depleted uranium (DU) pinhole collimators are evaluated for small-volume SPECT imaging of 511-keV gammas, with a view toward small-animal imaging. Tungsten (W) pinholes are examined for comparison. Line source measurements show planar full-width at half-maximum (FWHM) as low as 6.5 mm for a 1-mm DU pinhole, with SPECT system FWHM of 5.0 mm at 43-mm ROR and 6.7 mm at 69-mm ROR. Planar FWHM is smaller for DU than for W pinholes but is insensitive to aperture size for pinholes of 1-3 mm diameter at this energy for the collimators studied. These trends indicate penetration at pinhole edges, as confirmed by LSF shape and full-width at tenth-maximum trends. In SPECT scans of rod phantoms at 71-mm ROR, DU pinholes were only barely able to resolve cold rods as large as 4.7 mm (center-to-center spacing=2×diameter). For both F-18 alone and F-18 imaged along with Tc-99m, 3.9-mm hot rods were resolved. Tc-99m scans resolved 2.3- mm rods with F-18 present and 1.5-mm rods without F-18. DU pinhole SPECT at 511 keV does not currently match the spatial resolution of high-resolution positron emission tomography (PET) but is open to improvement and, unlike PET, is capable of simultaneous multiple-isotope acquisitions
  • Keywords
    biomedical equipment; single photon emission computed tomography; uranium; 1 to 3 mm; 511 keV; 511-keV photon SPECT imaging; F; Tc; U; W; aperture size; cold rods; depleted uranium pinhole collimators; pinhole edges; rod phantoms; small volumes; small-animal imaging; Apertures; Energy resolution; Imaging phantoms; Optical collimators; Optical imaging; Positron emission tomography; Shape; Single photon emission computed tomography; Spatial resolution; Tungsten;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.958384
  • Filename
    958384