    Calibration system of the Phobos silicon detectors

  • Author

    Griesmayer, Erich ; Frais-Kölbl, Helmut ; Pernegger, Heinz ; Wadsworth, Bernard ; Fitch, John

  • Author_Institution
    Fachhochschule Wiener Neustadt, Austria
  • Volume
    48
  • Issue
    4
  • fYear
    2001
  • fDate
    8/1/2001 12:00:00 AM
  • Firstpage
    1565
  • Lastpage
    1569
  • Abstract
    The Phobos experiment at the Relativistic Heavy Ion Collider (RHIC) at Brookhaven National Laboratory uses silicon pad detectors to measure charged particles in Au-Au collisions. Precise knowledge of the ionization signal amplitude in the silicon pad detector is vital for the determination of particle multiplicity and for particle identification. We describe here a system that precisely calibrates the gain curve for each channel in the 135 000-channel silicon readout system. The calibration covers the range up to 100 minimum ionizing particles, and the calibration circuit has a linearity better than ±1% over a dynamic range of 1:250
  • Keywords
    calibration; nuclear electronics; readout electronics; silicon radiation detectors; Phobos silicon detectors; RHIC; Relativistic Heavy Ion Collider; Si; calibration circuit; calibration system; ionization signal amplitude; minimum ionizing particles; particle identification; particle multiplicity; Calibration; Circuits; Current measurement; Detectors; Ionization; Laboratories; Linearity; Particle measurements; Signal processing; Silicon;