A simple method of energy calibration for thin plastic scintillator
Author
Hohara, S. ; Saiho, F. ; Tanaka, J. ; Aoki, S. ; Uozumi, Y. ; Matoba, M.
Author_Institution
Graduate Sch. of Eng., Kyushu Univ., Fukuoka, Japan
Volume
48
Issue
4
fYear
2001
fDate
8/1/2001 12:00:00 AM
Firstpage
1172
Lastpage
1176
Abstract
A thin plastic scintillator is used as a trigger counter in a charged particle spectrometer. Since its Compton edge is not clear, it is difficult to test the thin plastic scintillator counter with a standard gamma-ray source. In this paper, a simple method is proposed to determine the Compton edge and the energy resolution of the thin plastic scintillator by fitting the simulated spectrum to the measured spectrum. The simulated spectrum was made with a Monte Carlo program utilizing the geometry of the plastic scintillator. This technique was examined with three different plastic scintillators
Keywords
calibration; gamma-ray detection; solid scintillation detectors; Compton edge; Monte Carlo program; charged particle spectrometer; energy calibration; energy resolution; standard gamma-ray source; thin plastic scintillator; trigger counter; Calibration; Energy measurement; Energy resolution; Geometry; Monte Carlo methods; Plastics; Radiation detectors; Solid modeling; Spectroscopy; Testing;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/23.958745