• DocumentCode
    1542788
  • Title

    SQUID gradiometer measurement system for magnetorelaxometry in a disturbed environment

  • Author

    Schambach, J. ; Warzemann, L. ; Weber, P. ; Kotitz, R. ; Weitschies, W.

  • Author_Institution
    Friedrich-Schiller-Univ., Jena, Germany
  • Volume
    9
  • Issue
    2
  • fYear
    1999
  • fDate
    6/1/1999 12:00:00 AM
  • Firstpage
    3527
  • Lastpage
    3530
  • Abstract
    A SQUID measurement system for magneto-relaxometry (MRX) in disturbed environment is presented. It was developed for the determination of the distribution of magnetic nanoparticles in large objects like animals. The system is based on a thin film LTS SQUID gradiometer with a baseline of 5 mm and a field gradient sensitivity of about 30 fT/cmHz/sup 1/2/. The entire system is PC-controlled and includes a nonmagnetic x-y stage. A Helmholtz coil of 80 cm diameter allows the magnetization of the samples with fields up to 7 mT. First measurements in a disturbed environment are presented.
  • Keywords
    SQUID magnetometers; magnetic particles; magnetic relaxation; nanostructured materials; 7 mT; Helmholtz coil; PC control; animal; disturbed environment; magnetic nanoparticle; magnetization; magnetorelaxometry; measurement system; thin film LTS SQUID gradiometer; Coils; Immune system; Magnetic domains; Magnetic field measurement; Magnetic flux; Magnetic separation; Magnetic shielding; Magnetization; Nanoparticles; SQUIDs;
  • fLanguage
    English
  • Journal_Title
    Applied Superconductivity, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    1051-8223
  • Type

    jour

  • DOI
    10.1109/77.783791
  • Filename
    783791