For pt.I see ibid., vol.44, pp.259-70 (1997). The performance of the maximum likelihood closed loop circuits, proposed in part I for phase aberration correction in phased array imaging systems, is analysed. This analysis is helpful in designing and also in determining the tracking mode phase error variance performance of both closed loop circuits (1-D and 2-D). It is an approximate analysis, suitable under small errors conditions, and considered accurate for high values of the signal to noise ratio.
Keywords
aberrations; acoustic signal processing; biomedical ultrasonics; closed loop systems; maximum likelihood estimation; medical signal processing; ultrasonic transducer arrays; 1-D; 2-D; closed loop ML algorithm; closed loop circuits; maximum likelihood closed loop circuits; performance analysis; phase aberration correction; phased array imaging systems; signal to noise ratio; small errors conditions; tracking mode phase error variance performance; ultrasound medical imaging; Analysis of variance; Circuits; Maximum likelihood estimation; Performance analysis; Phase estimation; Phase measurement; Phased arrays; Signal processing algorithms; Signal to noise ratio; Tracking loops;
fLanguage
English
Journal_Title
Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
Publisher
ieee
ISSN
0885-3010
Type
jour
DOI
10.1109/58.585112
Filename
585112
Link To Document