DocumentCode
1549281
Title
Computer control of a slow-AC technique for thermopower measurements
Author
Marcy, Henry O. ; Marks, Tobin J. ; Kannewurf, Carl R.
Author_Institution
Northwestern Univ., Evanston, IL, USA
Volume
39
Issue
5
fYear
1990
fDate
10/1/1990 12:00:00 AM
Firstpage
756
Lastpage
760
Abstract
A measurement algorithm and the required instrumentation for computer automation of temperature-dependent thermopower measurements over the range of 4.2 to 350 K are described. The performance of the measurement system is demonstrated by examining typical results obtained on bulk and thin film samples of the thallium-based ceramic superconductor. The quality of the measurement data obtained is improved by increasing the density of the collected data and by improving reproducibility of the results
Keywords
barium compounds; calcium compounds; computerised control; computerised instrumentation; electric variables measurement; high-temperature superconductors; thallium compounds; thermoelectricity; 4.2 to 350 K; Tl-Ba-Ca-Cu-O; bulk simple; ceramic superconductor; computer automation; computerised instrumentation; data acquisition; high temperature superconductor; reproducibility; thermoelectric power; thermopower; thin film samples; Automatic control; Automation; Charge measurement; Conductivity measurement; Current measurement; Electrical resistance measurement; Temperature dependence; Temperature distribution; Temperature measurement; Voltage;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/19.58621
Filename
58621
Link To Document