Title :
Implantable defibrillator with high-output pacing function after defibrillation
Author :
Makino, Hideo ; Saitoh, Yoshiaki ; Mitamura, Yoshinori ; Mikami, Tomohisa
Author_Institution :
Fac. of Eng., Niigata Univ., Japan
Volume :
76
Issue :
9
fYear :
1988
fDate :
9/1/1988 12:00:00 AM
Firstpage :
1187
Lastpage :
1193
Abstract :
An experimental implantable defibrillator with a temporary pacing function was developed. The pacing function was activated after defibrillation to work as a high-output ventricular-demand pacemaker. The developed system was tested in animal experiments involving anesthesized and mechanically ventilated mongrel dogs. In the animal experiments, fibrillation was electrically induced, and the automatic defibrillation and pacing function were confirmed. After defibrillation, it was possible to continue the fixed-rate pacing for more than three minutes
Keywords :
defibrillators; experimental implantable defibrillator; high-output pacing function; high-output ventricular-demand pacemaker; mechanically ventilated mongrel dogs; Animals; Circuits; Defibrillation; Electrets; Electrocardiography; Electrodes; Fibrillation; Heart beat; Pacemakers; Pulse generation;