DocumentCode :
1552064
Title :
Automated Prescription of an Optimal Imaging Plane for Measurement of Cerebral Blood Flow by Phase Contrast Magnetic Resonance Imaging
Author :
Teng, Pang-yu ; Bagci, Ahmet Murat ; Alperin, Noam
Author_Institution :
Dept. of Bioeng., Univ. of Illinois at Chicago, Chicago, IL, USA
Volume :
58
Issue :
9
fYear :
2011
Firstpage :
2566
Lastpage :
2573
Abstract :
This study describes and evaluates a semiautomated method for prescribing an optimal imaging plane that is located as close as possible to the skull base, and is simultaneously nearly perpendicular to the four arteries leading blood to the brain [internal carotid arteries (ICAs) and vertebral arteries (VAs)]. Such a method will streamline and improve reliability of the measurement of total cerebral blood flow and intracranial pressure by velocity encoding phase-contrast magnetic resonance imaging. The method first extracts the vessels´ centerline from a 2-D time-of-flight magnetic resonance angiogram of the neck by performing distance transformations. An anatomical marker, the V2 segment of the VAs, is then identified to guide the imaging plane to be as close and below the skull base. An imaging plane that is nearly perpendicular to the ICAs and V2 segment of VAs is then identified by minimizing a misalignment value, estimated by a weighted mean of the angles between the plane´s normal and the vessel axes at the vessel-plane intersections. The performance of the semiautomated method was evaluated by comparing manually selected planes to those found semiautomatically in nine magnetic resonance angiogram datasets. The semiautomated method consistently outperformed manual prescription with a significantly smaller misalignment value, 8.6° versus 20.7° (P <; 0.001), respectively, and significantly improved reproducibility.
Keywords :
biomedical MRI; blood flow measurement; blood vessels; brain; data analysis; 2D time-of-flight magnetic resonance angiogram; ICAs; V2 segment; anatomical marker; cerebral blood flow measurement; internal carotid arteries; intracranial pressure measurement; magnetic resonance angiogram datasets; neck; optimal imaging plane; semiautomated method; skull base; velocity encoding phase-contrast magnetic resonance imaging; vertebral arteries; vessel centerline; vessel-plane intersections; Blood flow; Fluid flow measurement; Image segmentation; Magnetic resonance imaging; Manuals; Skeleton; Computer-guided procedure; magnetic resonance scan prescription; phase contrast magnetic resonance imaging; total cerebral blood flow quantitation; vessel centerline identification; Adult; Brain; Carotid Artery, Internal; Cerebrovascular Circulation; Databases, Factual; Female; Humans; Image Processing, Computer-Assisted; Magnetic Resonance Angiography; Male; Middle Aged; Vertebral Artery;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/TBME.2011.2159383
Filename :
5873130
Link To Document :
بازگشت