    Implementation of the beam reversal technique on compact cesium clocks: towards an improvement in accuracy

  • Author

    Chassagne, Luc ; Hamouda, Frederic ; Théobald, Genevieve ; Crez, P.

  • Author_Institution
    Lab. de l´´Horloge Atomique, CNRS, Orsay, France
  • Volume
    48
  • Issue
    6
  • fYear
    2001
  • Firstpage
    1513
  • Lastpage
    1516
  • Abstract
    Reports the evaluation of the residual phase difference /spl utri//spl phi/ in a short (18 cm) Ramsey cavity by implementing the beam reversal technique in an optically pumped cesium beam clock. /spl utri//spl phi/ is measured to be 21 /spl plusmn/1.5 /spl mu/rad, allowing a more accurate evaluation of the frequency performance of this small cesium clock. Finally, the clock accuracy is equal to 1.1/spl middot/10/sup -13/.
  • Keywords
    atomic clocks; caesium; cavity resonators; frequency standards; optical pumping; 18 cm; Cs; beam reversal technique; clock accuracy; compact cesium clocks; frequency performances; optically pumped clock; residual phase difference; short Ramsey cavity; Clocks; Frequency measurement; Laboratories; Laser beams; Laser excitation; Microwave magnetics; Optical beams; Optical pumping; Optical resonators; Ovens;