Title :
The liquid xenon detector for PET: recent results
Author :
Chepel, V. ; Solovov, V. ; Van Der Marel, J. ; Lopes, M.I. ; Crespo, P. ; Janeiro, L. ; Santos, Dinis ; Marques, R. Ferreira ; Policarpo, A.J.P.L.
Author_Institution :
Dept. de Fisica, Coimbra Univ., Portugal
fDate :
8/1/1999 12:00:00 AM
Abstract :
A multiwire ionisation chamber filled with liquid xenon as a position sensitive detector for Positron Emission Tomography is studied. Both xenon scintillation, which provides a fast trigger, and ionisation charge for position and energy measurements are read. The detection efficiency of the detector is assessed in separate measurements of the probability of detection of the scintillation light and the charge signals due to 511 keV gamma-rays. These were measured to be about 1 and 0.82, respectively
Keywords :
gamma-ray detection; liquid ionisation chambers; positron emission tomography; xenon; 511 keV; PET; Xe; charge signals; energy measurements; fast trigger; ionisation charge; liquid xenon detector; medical diagnostic imaging; medical instrumentation; multiwire ionisation chamber; nuclear medicine; position measurements; position sensitive detector; scintillation light detection; xenon scintillation; Charge measurement; Current measurement; Energy measurement; Gamma ray detection; Gamma ray detectors; Ionization chambers; Position sensitive particle detectors; Positron emission tomography; Solid scintillation detectors; Xenon;
Journal_Title :
Nuclear Science, IEEE Transactions on