DocumentCode
1555
Title
A High Count Rate Neutron Beam Monitor for Neutron Scattering Facilities
Author
Barnett, A.K. ; Cox, M.N. ; Crow, Lowell ; Diawara, Y. ; Funk, L.L. ; Hayward, J.P. ; Menhard, K. ; Sedov, V.N.
Author_Institution
Univ. of Tennessee, Knoxville, TN, USA
Volume
60
Issue
2
fYear
2013
fDate
Apr-13
Firstpage
668
Lastpage
670
Abstract
Beam monitors are an important diagnostic tool in neutron science facilities. Present beam monitors use either ionization chambers in integration mode, which are slow and have no timing information, or pulse counters which can easily be saturated by high beam intensities. At high flux neutron scattering facilities, neutron beam monitors with very low intrinsic efficiency (10-5) are presently selected to keep the counting rate within a feasible range, even when a higher efficiency would improve the counting statistics and yield a better measurement of the incident beam. In this work, we report on a high count rate neutron beam monitor. This beam monitor offers good timing with an intrinsic efficiency of 10-3 and a counting rate capability of over 1,000,000 cps without saturation.
Keywords
ionisation chambers; particle beam diagnostics; beam intensities; diagnostic tool; flux neutron scattering facilities; high count rate neutron beam monitor; integration mode; ionization chambers; neutron scattering facilities; neutron science facilities; pulse counters; Aluminum; Laboratories; Monitoring; Neutron spin echo; Neutrons; Particle beams; Radiation detectors; Gas detectors; neutron detectors;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.2012.2225111
Filename
6407474
Link To Document