DocumentCode :
1558690
Title :
Unified deadtime correction model for PET
Author :
Daube-Witherspoon, Margaret E. ; Carson, Richard E.
Author_Institution :
Dept. of Nucl. Med., Nat. Inst. of Health, Bethesda, MD, USA
Volume :
10
Issue :
3
fYear :
1991
fDate :
9/1/1991 12:00:00 AM
Firstpage :
267
Lastpage :
275
Abstract :
A model of deadtime for emission and transmission scans in positron emission tomography (PET) scanners with two-dimensional detectors has been developed. The model takes into account coincidence losses due to singles losses and multiple events, as well as mispositioning errors at higher count rates caused by pulse pile-up, within a detector block. The model is applicable to emission distributions and to spatially varying singles distributions seen with a rotating pin transmission source. An automatic procedure to determine the parameters of this model based on decaying emission studies has also been developed. Different singles dead time factors are required for emission and blank distributions due to differences in their energy spectra. The model was tested on emission and pin transmission data taken on the Scanditronix PC2048-15B scanner
Keywords :
computerised tomography; modelling; radioisotope scanning and imaging; Scanditronix PC2048-15B scanner; automatic procedure; coincidence losses; detector block; emission scans; energy spectra; medical diagnostic imaging; mispositioning errors; multiple events; nuclear medicine; positron emission tomography; pulse pile-up; rotating pin transmission source; singles losses; transmission scans; unified deadtime correction model; Attenuation; Calibration; Cities and towns; Contamination; Detectors; Event detection; Nuclear and plasma sciences; Nuclear medicine; Positron emission tomography; Testing;
fLanguage :
English
Journal_Title :
Medical Imaging, IEEE Transactions on
Publisher :
ieee
ISSN :
0278-0062
Type :
jour
DOI :
10.1109/42.97575
Filename :
97575
Link To Document :
بازگشت