DocumentCode :
1558693
Title :
Use of transputers in a 3-D positron emission tomograph
Author :
Atkins, M. Stella ; Murray, Donald ; Harrop, Ronald
Author_Institution :
Sch. of Comput. Sci., Simon Fraser Univ., Burnaby, BC, Canada
Volume :
10
Issue :
3
fYear :
1991
fDate :
9/1/1991 12:00:00 AM
Firstpage :
276
Lastpage :
283
Abstract :
The use of a VME-bus-based transputer network as a parallel processing engine for positron volume imaging (PVI) is discussed. The authors find that the speedups of parallel networks depend on two major factors, the ratio of computation to communication for a task and the size of the task, and give a simple model to explore the limits on speedups. Through actual implementation it is shown that real-time PVI data acquisition can be achieved with about 20 transputer nodes, and it is estimated that three-dimensional (3-D) image reconstruction can be achieved within 10 min using 200 nodes. Larger images and a larger number of histograms can readily be accommodated using the same parallel algorithms, as the model presented places no limits on the size of the images. The versatility and scalability of transputers makes them very suitable for use in PVI tomographs in that the same transputers can be used for speeding up data acquisition, image reconstruction, and display
Keywords :
computerised tomography; medical diagnostic computing; radioisotope scanning and imaging; transputers; 10 min; 3D image reconstruction; 3D positron emission tomograph; VME-bus-based transputer network; data acquisition; histograms; medical diagnostic imaging; nuclear medicine; parallel algorithms; parallel processing engine; positron volume imaging; Computer networks; Concurrent computing; Data acquisition; Engines; Histograms; Image reconstruction; Parallel algorithms; Parallel processing; Positrons; Radioactive decay;
fLanguage :
English
Journal_Title :
Medical Imaging, IEEE Transactions on
Publisher :
ieee
ISSN :
0278-0062
Type :
jour
DOI :
10.1109/42.97576
Filename :
97576
Link To Document :
بازگشت