• DocumentCode
    1560379
  • Title

    Optimization of the temporal response of II-VI direct type semiconductor detectors for flat-panel pulsed X-ray imaging

  • Author

    Giakos, George C. ; Guntupalli, R. ; Nemer, R. ; Odogba, J. ; Shah, N. ; Vedantham, S. ; Suryanarayanan, S. ; Chowdhury, S. ; Passerini, A.G. ; Mehta, K. ; Sumrain, S. ; Patnekar, N. ; Nataraj, K. ; Evans, E. ; Russo, F.

  • Author_Institution
    Dept. of Electr. Eng., Akron Univ., OH, USA
  • Volume
    50
  • Issue
    6
  • fYear
    2001
  • fDate
    12/1/2001 12:00:00 AM
  • Firstpage
    1610
  • Lastpage
    1614
  • Abstract
    The rising and falling edges of detected signal pulses have been measured utilizing X-ray ionization of a planar Cd1-xZnx Te system under different irradiation geometries, at different detector thicknesses, and applied electric fields. The experimental results of this study indicate that the time response of the CdZnTe based X-ray system is suitable for digital pulsed radiographic applications
  • Keywords
    II-VI semiconductors; X-ray detection; X-ray imaging; cadmium compounds; diagnostic radiography; semiconductor devices; zinc compounds; CdZnTe; CdZnTe based X-ray system; II-VI direct type semiconductor detectors; X-ray ionization; applied electric fields; detector thicknesses; digital pulsed radiographic applications; flat-panel digital radiography; flat-panel pulsed X-ray imaging; irradiation geometries; planar Cd1-xZnxTe system; temporal response optimisation; time response; Electric variables measurement; Geometry; Ionization; Pulse measurements; Signal detection; Tellurium; Thickness measurement; X-ray detection; X-ray detectors; Zinc;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.982954
  • Filename