Title :
A new method of radio frequency links by coplanar coils for implantable medical devices
Author :
Xue, L. ; Hao, H.W. ; Li, L.M. ; Ma, B.Z.
Author_Institution :
Tsinghua Space Center, Tsinghua Univ., Beijing
Abstract :
A new method based on coplanar coils for the design of radio frequency links has been developed, to realize the communication between the programming wand and the implantable medical devices with shielding container simply and reliably. With the analysis of electronic and magnetic field theory, the communication model has been established and simulated, and the circuit has been designed and tested. The experimental results are consistent with the simulation fairly well. The voltage transfer ratio of the typical circuit with present parameters can reach as high as 0.02, which can fulfil the requirements of communication
Keywords :
biomedical equipment; coils; prosthetics; coplanar coils; electromagnetic shielding; electronic field theory; implantable medical devices; magnetic field theory; programming wand; radio frequency links; voltage transfer ratio; Analytical models; Circuit simulation; Circuit testing; Coils; Containers; Implantable biomedical devices; Magnetic analysis; Magnetic shielding; Medical simulation; Radio frequency; electromagnetic coupling; electromagnetic shield; implantable medical device; mutual inductance coils; radio frequency link;
Conference_Titel :
Engineering in Medicine and Biology Society, 2005. IEEE-EMBS 2005. 27th Annual International Conference of the
Conference_Location :
Shanghai
Print_ISBN :
0-7803-8741-4
DOI :
10.1109/IEMBS.2005.1615293