DocumentCode :
1571090
Title :
Differential ISFET/REFET pairs as a reference system for integrated ISFET-sensor arrays
Author :
Muller, E. ; Woias, P. ; Hein, P. ; Koch, S.
Author_Institution :
Tech. Univ., Munchen, Germany
fYear :
1991
Firstpage :
467
Lastpage :
470
Abstract :
In order to make use of all the advantages over conventional glass electrodes which an ISFET sensor system provides, an integrated reference electrode has to be developed. The concept of using a noble metal as a pseudo-reference electrode requires a different ISFET/REFET (ion-sensitive FET/reference FET) pair. Two types of reference system are compared. The first system is a pair of ISFETs, one with the pH sensitivity of a bare Si/sub 3/N/sub 4/ gate insulator (ca. 50 mV/pH) and a second with a reduced sensitivity in the range of 25 mV/pH. This pseudo-reference-FET was realized by depositing an extremely thin polymeric film on the gate insulator. An enzyme FET was studied as the second system to examine the application of a pseudo-reference electrode to biosensors. Both systems show similar behavior with regard to the influence of the pseudo-reference electrode. The results have led to the design of an ISFET sensor array which is optimized for using an integrated noble metal electrode as a solution contact.<>
Keywords :
CMOS integrated circuits; biosensors; electric sensing devices; electrochemical analysis; electrochemical electrodes; insulated gate field effect transistors; Si/sub 3/N/sub 4/ gate insulator; biosensors; design; differential ISFET/REFET pairs; enzyme FET; integrated ISFET-sensor arrays; integrated reference electrode; noble metal; pH sensitivity; pseudo-reference electrode; pseudo-reference-FET; reference FET; thin polymeric film; Biochemistry; Biosensors; Design optimization; Electrodes; FETs; Glass; Plastic insulation; Polymer films; Sensor arrays; Sensor systems;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Solid-State Sensors and Actuators, 1991. Digest of Technical Papers, TRANSDUCERS '91., 1991 International Conference on
Conference_Location :
San Francisco, CA, USA
Print_ISBN :
0-87942-585-7
Type :
conf
DOI :
10.1109/SENSOR.1991.148913
Filename :
148913
Link To Document :
بازگشت