• DocumentCode
    1575703
  • Title

    A new calibration method for magnetic sensor array for tracking capsule endoscope

  • Author

    Li, Mao ; Song, Shuang ; Hu, Chao ; Yang, Wanan ; Wang, Lujia ; Meng, Max Q H

  • Author_Institution
    Key Lab. for Biomed. Inf. & Health Eng., CAS, Shenzhen, China
  • fYear
    2009
  • Firstpage
    1561
  • Lastpage
    1566
  • Abstract
    To track the movement of wireless capsule endoscope in the human body, we design a magnetic localization and orientation system. In this system, capsule contains a permanent magnet as the movable object. A wearable magnetic sensor array is arranged out of the human body to capture the magnetic signal. This sensor array is composed of magnetic sensors, Honeywell product HMC1043. The variations of magnet field intensity and direction are related to the capsule position and orientation. Therefore, the 3D localization information and 2D orientation parameters of capsule can be computed based on the captured magnetic signals and by applying an appropriate algorithm. In order to initialize the system and improve the tracking accuracy, we propose a calibration technique based on high-accurate localization equipment, FASTRAK. The calibration method includes two steps. Firstly, we acquire the accurate reference data from FASTRAK tracking equipment, and transform them into the position and orientation parameters of the magnet. Secondly, we calculate three important parameters for the sensor calibration: the sensitivity, the center position, and the orientation. Based on the calibration, we can adjust the magnetic localization and orientation system quickly and accurately. The experimental results prove that the calibration method used in our system can improve the system with satisfactory tracking accuracy.
  • Keywords
    calibration; diseases; endoscopes; magnetic sensors; medical signal detection; numerical analysis; permanent magnets; tracking; 2D orientation parameters; 3D localization information; FASTRAK tracking equipment; Honeywell product HMC1043; calibration method; capsule endoscope tracking; capsule position; captured magnetic signals; diseases; high-accurate localization equipment; magnetic localization; magnetic sensor array; magnetic signal; orientation system; permanent magnet; sensor calibration; Calibration; Endoscopes; Humans; Magnetic devices; Magnetic sensors; Permanent magnets; Sensor arrays; Tracking; Wearable sensors; Wireless sensor networks;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Robotics and Biomimetics (ROBIO), 2009 IEEE International Conference on
  • Conference_Location
    Guilin
  • Print_ISBN
    978-1-4244-4774-9
  • Electronic_ISBN
    978-1-4244-4775-6
  • Type

    conf

  • DOI
    10.1109/ROBIO.2009.5420397
  • Filename
    5420397