  • Title

    A calorimeter for μW level optical power transfer standard

  • Author

    Suzuki, Yasuyuki ; Chida, Naomichi ; Komiyama, Makoto

  • Author_Institution
    Yokogawa Electr. Corp., Tokyo, Japan
  • fYear
    1991
  • Firstpage
    131
  • Lastpage
    134
  • Abstract
    A calorimeter than can measure μW-level optical power has been developed for an optical power transfer standard. It can measure the optical fiber simply by attaching an adapter. This calorimeter sensor uses a compensative absorber to reduce the influence of pressure fluctuation and temperature variation in the measurement room, and an ultra-low-noise preamplifier. With this calorimeter, the standard deviation of the measured value is in the range of 0.02% to 0.4% in 20-mW to 10-μW optical power measurements. An error evaluation for an optical power level of 10 μW yielded a two-sigma (two standard deviation) total uncertainty of 0.9%
  • Keywords
    calorimeters; measurement standards; optical fibres; optical variables measurement; power measurement; calorimeter sensor; compensative absorber; error evaluation; measurement standards; optical fiber; optical power measurements; optical power transfer standard; pressure fluctuation; standard deviation; Fluctuations; Joining processes; Measurement standards; Optical fibers; Optical sensors; Power measurement; Preamplifiers; Pressure measurement; Standards development; Temperature sensors;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Instrumentation and Measurement Technology Conference, 1991. IMTC-91. Conference Record., 8th IEEE
  • Conference_Location
    Atlanta, GA
  • Print_ISBN
    0-87942-579-2
  • Type

    conf

  • DOI
    10.1109/IMTC.1991.161558
  • Filename
    161558