DocumentCode
1580173
Title
Static error in the temperature constant measurement configurations due to DC amplifier offset voltage
Author
Freire, Raimundo C S ; Santos, Francisco F. ; Alves, Raimundo N C ; Barros, Alessio T. ; Luciano, Benedito A.
Author_Institution
Univ. Fed. da Campina Grande, Brazil
Volume
2
fYear
2003
Firstpage
1607
Abstract
The DC amplifier offset voltage introduces dynamic and static errors in measurements based on heated sensors at constant temperature configuration. It is shown, in the case of a radiometer, the variation of static errors in function of offset voltage for configurations of feedback based on analogic voltage control and on pulse width modulation - PWM. These errors increase with offset voltage and are similar in PWM configuration, decreasing with pulse amplitude.
Keywords
DC amplifiers; measurement errors; pulse width modulation; radiometers; temperature measurement; DC amplifier offset voltage; PWM configuration; analogic voltage control; heated sensors; pulse amplitude; pulse width modulation; radiometer; static error; temperature constant measurement; Error correction; Feedback; Pulse amplifiers; Pulse width modulation; Radiometry; Space vector pulse width modulation; Temperature measurement; Temperature sensors; Thermal sensors; Voltage;