• DocumentCode
    1580448
  • Title

    Improvements in ultrasound elastography using dynamic focusing

  • Author

    Lertsilp, D. ; Umchid, S. ; Techavipoo, U. ; Thajchayapong, P.

  • Author_Institution
    Dept. of Ind. Phys. & Med., King Mongkut´´s Univ. of Technol. North Bangkok, Bangkok, Thailand
  • fYear
    2012
  • Firstpage
    225
  • Lastpage
    228
  • Abstract
    Many ultrasound elastograms are reconstructed from pre- and post-compression radiofrequency (RF) signals after beamforming using a single or a few focusing depths. This might limit the accuracy of the displacement estimates for elastography in the area far from the focusing depths. This article studies the improvement of the elastograms reconstructed from the signals beamformed using a dynamic focusing scheme compared with a single focusing scheme. The results show that the elastograms reconstructed using the signals from the dynamic focusing scheme have higher signal-to-noise ratio (SNRe) and contrast-to-noise ratio (CNRe) than those from the single focusing scheme.
  • Keywords
    biomedical ultrasonics; medical signal processing; signal reconstruction; ultrasonic focusing; beamforming; contrast-to-noise ratio; dynamic focusing; signal reconstruction; signal-to-noise ratio; ultrasound elastography; CNRe; SNRe; dynamic focusing; single focusing; ultrasound elastography;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Biomedical Engineering International Conference (BMEiCON), 2011
  • Conference_Location
    Chiang Mai
  • Print_ISBN
    978-1-4577-2189-2
  • Type

    conf

  • DOI
    10.1109/BMEiCon.2012.6172057
  • Filename
    6172057