Title : 
Breathing rate estimation from a non-contact biosensor using an adaptive IIR notch filter
         
        
            Author : 
Ballal, Tarig ; Shouldice, Redmond B. ; Heneghan, Conor ; Zhu, Anding
         
        
            Author_Institution : 
Sch. of Electr., Electron. & Commun. Eng., Univ. Coll. Dublin, Dublin, Ireland
         
        
        
        
        
            Abstract : 
In this paper, a non-contact method for human breathing rate estimation is discussed. The method utilises SleepMinder technology, which implements a radio-frequency Doppler radar system to capture physiological movements of the body in the form of phase modulation. To determine the breathing rate, the signals are down-converted and the baseband signals are processed. Previously reported methods used the conventional periodogram approach to estimate the breathing rate. A downside of the periodogram method is its limited capability in capturing respiration dynamics. The adaptive notch filter approach, discussed herein, provides a better respiration rate tracking performance than the periodogram approach, as demonstrated by the experimental results presented in this paper.
         
        
            Keywords : 
Doppler radar; IIR filters; adaptive filters; biosensors; medical signal processing; notch filters; pneumodynamics; SleepMinder technology; adaptive IIR notch filter; baseband signal processing; conventional periodogram approach; down-converted processing; human breathing rate estimation; noncontact biosensor; physiological body movements; radiofrequency Doppler radar system; respiration dynamics; Baseband; Convergence; Doppler radar; Estimation; Frequency estimation; Least squares approximation; Breathing rate estimation; IIR notch filter; adaptive filter; frequency tracking; non-contact; non-invasive; periodogram;
         
        
        
        
            Conference_Titel : 
Biomedical Wireless Technologies, Networks, and Sensing Systems (BioWireleSS), 2012 IEEE Topical Conference on
         
        
            Conference_Location : 
Santa Clara, CA
         
        
            Print_ISBN : 
978-1-4577-1135-0
         
        
        
            DOI : 
10.1109/BioWireless.2012.6172727