• DocumentCode
    1585798
  • Title

    The Development and Evaluation of New Neck Coil for GE MR System

  • Author

    Li, Bin ; Jie, Wang

  • Author_Institution
    Dept. of Clinical Eng., Shanghai Sixth People´´s Hosp.
  • fYear
    2006
  • Firstpage
    6033
  • Lastpage
    6036
  • Abstract
    Due to the MRI signal is much low, the improvement of image quality becomes the important goals for MRI equipment R&D. The RF coil is one of the most important factors in detecting the MRI signal. A new neck coil is developed with four channels phase array design and evaluated. Compared with original coil, the image signal to noise ratio (SNR) and uniformity of new neck coil are improved evidently
  • Keywords
    arrays; biomedical MRI; coils; GE MR system; MRI equipment; RF coil; coil uniformity; image quality improvement; image signal-to-noise ratio; neck coil; phase array design; Clinical diagnosis; Coils; Hospitals; Image quality; Magnetic field measurement; Magnetic resonance imaging; Neck; Phased arrays; Radio frequency; Signal to noise ratio; evaluation; image quality; neck coil;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 2005. IEEE-EMBS 2005. 27th Annual International Conference of the
  • Conference_Location
    Shanghai
  • Print_ISBN
    0-7803-8741-4
  • Type

    conf

  • DOI
    10.1109/IEMBS.2005.1615867
  • Filename
    1615867