Abstract :
The compact neutron spectrometer project was initiated to develop a device able to study fast neutrons produced by fusion reactions in a tokamak. The goal is to provide a fully characterized BC501A detector using a digital acquisition system developed by ENEA. The pulse height resolution and the response matrix of the detector are determined using experimental data acquired at the PTB facility. To achieve this goal, methods of data analysis involving entropy maximization and new software were developed. The methods are described and the current results for the pulse height resolution are presented.
Keywords :
Tokamak devices; neutron spectrometers; plasma diagnostics; BC501A compact neutron spectrometer; BC501A detector; compact neutron spectrometer project; digital acquisition system; entropy maximization; fast neutrons; fusion diagnostics; fusion reactions; pulse height resolution; response function determination; tokamak; Cyclotrons; Detectors; Energy resolution; Entropy; Minimization; Neutrons; Software;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Advancements in Nuclear Instrumentation Measurement Methods and their Applications (ANIMMA), 2011 2nd International Conference on
Conference_Location :
Ghent
Print_ISBN :
978-1-4577-0925-8
Type :
conf
DOI :
10.1109/ANIMMA.2011.6172866
Filename :
6172866
Link To Document :
بازگشت