• DocumentCode
    1588931
  • Title

    Non-amplifying ECG signal acquisition system based on Σ-Δ A/D

  • Author

    Liang, Hu ; Yanfang, Zhang ; Lesheng, He ; Weilian, Wang

  • Author_Institution
    Sch. of Inf. Sci. & Technol., Yunnan Univ., Kunming, China
  • Volume
    3
  • fYear
    2011
  • Firstpage
    105
  • Lastpage
    108
  • Abstract
    ECG signal is a very weak signal acquired on the surface of body. In traditional ECG signal acquisition system, to get a high-SNR ECG signal, it must use all kinds of high-performance op-amp, which leads to the acquiring circuits complex and unstable. Furthermore, almost all op-amps more or less have nonlinear distortion and other disadvantages. In this paper, a non-amplifying method to acquire a high-SNR ECG signal was proposed. It is based on 24 bit Σ-Δ A/D embedded in a high performance mixture signal ARM processor-ADuC7060. The system has high SNR up to 74.8 dB, and low nonlinear distortion, original ECG can be acquired.
  • Keywords
    electrocardiography; medical signal detection; operational amplifiers; ARM processor-ADuC7060; body surface; high performance op-amp; nonamplifying ECG signal acquisition system; Digital filters; Electrocardiography; Instruments; Low pass filters; Nonlinear distortion; Signal to noise ratio; Σ-Δ A/D; Non-Amplifying ECG acquisition; high-SNR; low nonlinear distortion;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Electronic Measurement & Instruments (ICEMI), 2011 10th International Conference on
  • Conference_Location
    Chengdu
  • Print_ISBN
    978-1-4244-8158-3
  • Type

    conf

  • DOI
    10.1109/ICEMI.2011.6037865
  • Filename
    6037865