• DocumentCode
    1590214
  • Title

    ADC nonlinearities and harmonic distortion in FFT test

  • Author

    Bellan, D. ; Brandolini, A. ; Gandelli, A.

  • Author_Institution
    Dipt. di Elettronica, Politecnico di Milano, Italy
  • Volume
    2
  • fYear
    1998
  • Firstpage
    1233
  • Abstract
    This paper deals with the analysis of harmonic distortion in uniform and nonuniform analog-to-digital converters (ADC), when the test signal is a pure sine wave. Mathematical relationships between harmonic distortion, input signal parameters and ADC nonlinearity errors are derived Finally, numerical simulations of A/D processes validate the analytical results
  • Keywords
    analogue-digital conversion; circuit analysis computing; digital simulation; fast Fourier transforms; harmonic distortion; integrated circuit testing; A/D process simulation; ADC nonlinearities; FFT test; harmonic distortion; input signal parameters; nonlinearity errors; nonuniform convertors; numerical simulation; test signal; uniform convertors; Analog-digital conversion; Analytical models; Harmonic analysis; Harmonic distortion; Numerical simulation; Quantization; Shape; Signal analysis; Signal processing; Testing;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Instrumentation and Measurement Technology Conference, 1998. IMTC/98. Conference Proceedings. IEEE
  • Conference_Location
    St. Paul, MN
  • ISSN
    1091-5281
  • Print_ISBN
    0-7803-4797-8
  • Type

    conf

  • DOI
    10.1109/IMTC.1998.676921
  • Filename
    676921