• DocumentCode
    1591811
  • Title

    The Accuracy of Surface Measurement for Motion Analysis of Osteoporotic Thoracolumbar Spine

  • Author

    Yang, Zhengyi ; Griffith, James F. ; Leung, Pc ; Pope, Malcolm ; Sun, Lw ; Lee, Raymond

  • Author_Institution
    Dept. of Rehabilitation Sci., Hong Kong Polytech. Univ.
  • fYear
    2006
  • Firstpage
    6871
  • Lastpage
    6874
  • Abstract
    The surface method of measuring the movements of lower thoracic and lumbar spine with osteoporosis using Fastrakreg was examined for validity. There is an increasing awareness of the health risks of exposure to radiation associated with repeated radiographic assessment of spinal curvature and spinal movements. As a skin-surface measurement device, Fastrakreg was employed to study the effect of low bone mineral density on spine motion. However, the reliability and validity of data recorded has not been established. The purpose of this study was to develop a methodology to determine the accuracy of the surface measurement device when it is applied to the motion analysis of osteoporotic spine. The results obtained indicated that the rotation angle of lumbar and lower thoracic spine could be predicted with acceptable accuracy from the data collected from skin-mounted sensors, while the prediction error of translational movements were not acceptable
  • Keywords
    biomechanics; biomedical equipment; bone; densitometry; diagnostic radiography; diseases; position measurement; sensors; Fastrak; health risks; low bone mineral density; lower thoracic movements; motion analysis; osteoporosis; osteoporotic thoracolumbar spine; prediction error; repeated radiographic assessment; skin-mounted sensors; skin-surface measurement device; spinal curvature; spinal movements; translational movements; Biomedical imaging; Bones; Kinematics; Minerals; Motion analysis; Motion measurement; Osteoporosis; Radiography; Spine; Sun;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 2005. IEEE-EMBS 2005. 27th Annual International Conference of the
  • Conference_Location
    Shanghai
  • Print_ISBN
    0-7803-8741-4
  • Type

    conf

  • DOI
    10.1109/IEMBS.2005.1616084
  • Filename