Edge detection in quantitative coronary arteriography (QCA): the impact of image zoom on edge positioning, accuracy, precision and discrimination
Author :
Linderer, T. ; Wunderlich, W. ; Backs, B. ; Noering, J. ; Schroeder, R.
Author_Institution :
Dept. of Cardiology, Klinikum Steglitz, Berlin, Germany
fYear :
1992
Firstpage :
99
Lastpage :
102
Abstract :
In geometric QCA, the edge detection (ED) overestimates vessels in the small diameter range (SR, <1.5 mm) due to imaging limitations. Subsequent diameter correction enforces accurate but imprecise corrected diameters. To improve the starting conditions for further analysis the authors used 1-6-fold zoom in phantoms and measured each diameter 240 times by automated ED with optimized weighting of the first and second derivative and subsequent isocenter calibration. The positioning of the second derivative is zoom dependent. In derivative weighting ED, zoom shifts the edges over the entire range toward the vessel center, while in first derivative based ED this occurs only in SR. Zoom improves the accuracy but deteriorates precision of the native measurements. Based on discrimination as the optimality criterion, 3-4-fold zoom is optimal. By applying optimal zoom, precision after diameter correction was doubled
Keywords :
cardiology; diagnostic radiography; medical image processing; derivative weighting; edge detection; edge positioning; image zoom; optimality criterion; phantoms; quantitative coronary arteriography; Angiography; Calibration; Cardiology; Displays; Image edge detection; Imaging phantoms; Noise measurement; Optical imaging; Quantum cellular automata; Strontium;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Computers in Cardiology 1992, Proceedings of
Conference_Location :
Durham, NC
Print_ISBN :
0-8186-3552-5
Type :
conf
DOI :
10.1109/CIC.1992.269437
Filename :
269437
Link To Document :
بازگشت