Accuracy and reproducibility of an on-line quantitative coronary angiographic system
Author :
Desmet, W. ; Willems, J.L. ; Vrolix, M. ; Van Lierde, J. ; Byttebier, G. ; Piessens, J.
Author_Institution :
Univ. Hospital Gasthuisberg, Leuven, Belgium
fYear :
1992
Firstpage :
539
Lastpage :
542
Abstract :
To evaluate the intraobserver and interobserver variability of an online quantitative coronary angiographic system, 166 lesions were measured three times by each of two independent observers before and after angioplasty. Before angioplasty 95% of the values for percent stenosis obtained by observers 1 and 2 were within a range of 13.2% and 15.5%, respectively, but these respective ranges increased to 19.2% and 16.8% after angioplasty. For measurements of absolute diameters of the minimal luminal diameter 95% limits of agreement between both observers were -0.26 to 0.28 before and -0.38 to 0.37 after PTCA. These ranges were independent of stenosis severity or vessel size
Keywords :
cardiology; diagnostic radiography; interobserver variability; intraobserver variability; medical diagnostic imaging; minimal luminal diameter; on-line quantitative coronary angiographic system; percent stenosis; stenosis severity; vessel size; Angiography; Angioplasty; Calibration; Digital images; Hospitals; Image analysis; Lesions; Pixel; Quantum cellular automata; Reproducibility of results;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Computers in Cardiology 1992, Proceedings of
Conference_Location :
Durham, NC
Print_ISBN :
0-8186-3552-5
Type :
conf
DOI :
10.1109/CIC.1992.269532
Filename :
269532
Link To Document :
بازگشت