DocumentCode
1595228
Title
The analysis of electrode-recording-horizon in multi-electrode array(MEA)
Author
Lin, Yunsheng ; Chen, Chuanping ; Chen, Lin ; Zeng, Shaoqun ; Luo, Qingming
Author_Institution
Hubei Bioinformatics & Molecular Imaging Key Lab., Huazhong Univ. of Sci. & Technol., Wuhan
fYear
2006
Firstpage
7345
Lastpage
7348
Abstract
There is a problem that can´t be ignored in the MEA collected-signal-sorting process: When a neuron positions in two adjacent microelectrodes, can its activity be detected at the same time by both microelectrodes? Under certain conditions, the contact between the electrode and the cultured cell could be simplified as capacitive contact. Because the distance and the covering area affect the capacitance, the amplitude of potential decreases rapidly with the increase of distance to the microelectrode. We show that common MEA chip whose spacing of electrodes is 200 mum can´t detect the neuronal potential in its adjacent electrodes simultaneously. About 100-recorded experiments data in our lab confirm this conclusion
Keywords
arrays; bioelectric potentials; capacitance; cellular biophysics; microelectrodes; neurophysiology; 200 mum; capacitance; cultured cell; electrode-recording-horizon; microelectrodes; multielectrode array; neuronal potential; signal-sorting process; Capacitance; Cells (biology); Crosstalk; Electrodes; Microelectrodes; Neurons; Platinum; Signal detection; Silicon; Sorting;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 2005. IEEE-EMBS 2005. 27th Annual International Conference of the
Conference_Location
Shanghai
Print_ISBN
0-7803-8741-4
Type
conf
DOI
10.1109/IEMBS.2005.1616208